5234. Clinical thermometers. (F. D. C No. 39450. S. Nos. 48-862/3 M.) QUANTITY : 28 doz. oral thermometers and 19 doz. rectal thermometers at Mil- waukee, Wis. SHIPPED : Between 3-24-56 and 7-3-56, from New York, N. Y., by the Philbern Thermometer Co., Inc. LABEL IN PAET: (Otn.) "One Fever Thermometer * * * Philbern * * * Oral (Or Rectal)." ACCOMPANYING LABELING: (Leaflet) "Certificate of Examination * * * The Government Bureau recognizes as reliable all thermometers having a correction of ?o of a degree at any test point * * * This clinical thermometer has been tested by a Standard Thermometer, approved by the Bureau of Standards, Washington, D. C, at the following degrees and found to stand at: 98?-0, 100?-0,104?-0,106?-0. If reading is plus then subtract from reading, if minus, add to reading." RESULTS OF INVESTIGATION : Examination of 24 oral thermometers and 24 rectal thermometers showed that 6 oral thermometers and 7 rectal thermometers failed to meet the labeled standard of accuracy. LIBELED : 8-29-56, E. Dist. Wis. CHARGE: 501 (c)-the quality of the thermometers, when shipped, fell below that which they purported and were represented to possess since their actual temperature readings did not comply with the standard of accuracy claimed for them. DISPOSITION : 10-11-56. Default-destruction. DRUGS AND DEVICES ACTIONABLE BECAUSE OF FALSE AND MISLEADING CLAIMS DRUGS FOR HUMAN USE*