5879. Rectal thermometers. (F.D.C. No. 42539. S. No. 32-231 P.) QUANTITY : 264 rectal thermometers in envelopes at Englewood, N.J. SHIPPED : 10-7-5S, from Brooklyn, N.Y., by Adelphi Surgical Co., Inc. LABEL IN PART: (Envelope) "Certificate of Examination Fever Thermometer * * * issued by the United States Dept. of Commerce Comet Thermometer Co., Brooklyn, N.Y. [or Cardinal Thermometer Co., Brooklyn, N.Y.]" and (engraved on thermometer) "Adeleo Rectal." RESULTS OF INVESTIGATION: Examination of 24 thermometers showed that 13 failed to meet the requirement for accuracy specified in CSl-52, issued by the National Bureau of Standards, when tested as described in CSl-52. LIBELED: 12-5-58, Dist. N.J. CHARGE: 501(c)-when shipped, the quality of the article fell below that which it purported and'was represented to possess since it did not give accurate readings; and 502(a)-the label statement "tested and found to meet all the requirements and tests specified in the Commercial Standard CSl-52" was false and misleading. DISPOSITION : 1-12-59. Default-destruction.